The HVC80502X X-ray source is an integrated X-ray generator with self-cooling and self-protecting features, with an output of up to 80kV/150W. It is characterized by its compact structure, portable installation, and reliable operation. Within the allowable working temperature range, it can operate continuously at 150W without additional cooling. The HVC80502X X-ray source consists of an X-ray chamber and a control box, and remote control, status monitoring, and firmware upgrades can be achieved via a standard RS232 interface.
Food testing, industrial non-destructive testing, dangerous goods testing and other fields, mostly used for simple X-ray machines or mobile X-ray testing equipment
Item | Specification |
Input voltage | 230VACยฑ10%, 50/60Hz, 2Amps |
Output power of X ray tube | Max continuous output power 150W๏ผ30kV/5mA or 80kV/1.8mA ๏ผ |
Output voltage | Rated output voltage๏ผContinuously adjustable voltage range 20kV—80kV |
Output voltage ripple๏ผยฑ0.5%๏ผpeak to peak๏ผ | |
Output voltage accuracy๏ผยฑ1% of voltage setting value | |
line regulation๏ผยฑ0.1% | |
load regulation๏ผยฑ0.1% | |
Tube current | Rated tube current๏ผContinuously adjustable current range 0.2mA-5.0mA |
Tube current accuracy๏ผยฑ1% mA of current setting value | |
line regulation๏ผยฑ0.5% | |
load regulation๏ผยฑ0.5% | |
Rise time of output voltage | The kV rise time is <0.4 Sec from 10% to 90% of the output voltage. |
Filament power supply: | input voltage๏ผ24VDC |
filament voltage๏ผ2.0 to 3.0Vac | |
filament๏ผ3.0 to 3.5 Amps RMS | |
preheating time๏ผ3sec | |
Tube feature | Tube type๏ผfixed anodeใ glass envelope ใtungsten target |
focus๏ผ0.8mm | |
inherent filtration๏ผ0.8mm Be๏ผ0.7mm AI | |
radiation angle๏ผ80ยบ*16ยบ, fan beam | |
target angle๏ผ25ยฐ | |
Cooling | Transformer oil medium, natural heat dissipation |
Working temperatures | -10โ—40โ |
Storing temperature | -20โ—60โ |
System temperature protection | 60ยฑ3โ of oil temperature |
Humidness | 98%๏ผNon-condensation |
Weight | 23kg |
Installation direction | Installation in any direction |
Radiation angle | 16ยฐร80ยฐ |
X-ray leakage | Less than 0.5mR/hr at 5cm from the surface of the HVC80502 Integrated X-Ray Source. |
JB1/AC~(AC Input Power Connector)
Pin | Signal | Parameter |
1 | L | live wire |
2 | N | Neutral line |
3 | G | PE |
JB2/COM (DMR-9S interface definitions)
Pin | Signal | Parameter |
1.4.6.7.8.9 | N/C | No connect |
2 | TXD | Data transmit |
3 | RXD | Data receive |
5 | GND | Signal gnd |
JB3/Interlock,(DMR-9P interface definitions)
Pin | Signal | Parameter |
3/4/5/6/7/8/9 | N/C | No connect |
1 | Interlock Out | |
2 | Interlock In |
Short connect pin1 and pin2 make X ray source normal operation. Typical connection๏ผ
Led indicator:
ID | Color | Meaning |
XrayOn | Yellow | indicate X ray on |
ARC | Red | Arcing in oil tank |
OT | Red | Over temperature |
EP_Err | Red | Tube voltage error |
IP_Err | Red | Tube current error |
Power | Green | Power on |
Tank size