HVC806A0 X-Ray Source

Key Features:

  • Integrated design with high electrical integration and a compact appearance
  • Capable of continuous, long-term operation
  • High stability
  • Versatile installation options
  • Standard digital interface for easy use

Product Details

The HVC806A0 is a compact, safe, and electrically stable integrated X-ray source with low leakage. It operates continuously at 240W within the allowable temperature range. The system includes a high-voltage power supply, filament power supply, X-ray tube, high-voltage oil tank, and radiator. This high-frequency, self-cooling, and self-protecting X-ray generator delivers up to a maximum voltage of up to 80kV and a maximum power of 240W. It mainly consists of a control box and high-voltage oil tank, with RS232 communication for control, monitoring, and firmware upgrades.

Application:

Food testing, industrial non-destructive testing, dangerous goods testing and other fields, mostly used for simple X-ray machines or mobile X-ray testing equipment

Specification:

Item Specification
Input voltage 230VACยฑ10%, 50/60Hz, 3Amps
Output power of X ray tube Max continuous output power 240W๏ผˆ80kV/3mA,40kv/6mA๏ผ‰
Output voltage Rated output voltage๏ผšContinuously adjustable voltage range 30kV—80kV
Output voltage ripple๏ผšยฑ0.5%๏ผˆpeak to peak๏ผ‰
Output voltage accuracy๏ผšยฑ2% of voltage setting value
line regulation๏ผšยฑ0.1%
load regulation๏ผšยฑ0.1%
Tube current Rated tube current๏ผšContinuously adjustable current range 1mA-6mA
Tube current accuracy๏ผšยฑ0.2mA of current setting value
line regulation๏ผšยฑ0.1%
load regulation๏ผšยฑ0.1%
Rise time of output voltage The kV rise time is <0.6 Sec at maximum power
The kV rise time is <0.1 Sec at low voltage ( < 40kV )
Filament power supply: input voltage๏ผš24VDC
filament voltage๏ผš2.0 to 5.3V
filament๏ผš3.0 to 3.8Amps RMS
preheating time๏ผš3sec
Tube feature Tube type๏ผšfixed anodeใ€glass envelopeใ€tungsten target
focus๏ผš0.8mm
inherent filtration๏ผš0.8mm Be๏ผŒ1.5mm AI
radiation angle๏ผš80ยบ*16ยบfan beam
target angle๏ผš25ยฐ
Cooling transformer oil๏ผŒexternal circulation๏ผŒexternal forced air cooling
Working temperatures -10โ„ƒ—40โ„ƒ
Storing temperature -20โ„ƒ—60โ„ƒ
System temperature protection 60โ„ƒยฑ3โ„ƒ of oil temperature
Humidness 98%๏ผŒNon- condensation
Weight 31.1kg
Installation direction Installation in any direction
Radiation angle 16ยฐร—80ยฐ
X-ray leakage Less than 0.5mR/hr at 5cm from the surface of the HVC806A0.

JB1/AC~(AC Input Power Connector)

Pin Signal Parameter
1 L live wire
2 N Neutral line
3 G PE

JB2/COM (DMR-9S interface definitions)

Pin Signal Parameter
1.4.6.7.8.9 N/C No connect
2 TXD Data transmit
3 RXD Data receive
5 GND Signal gnd

JB3/Interlock,(DMR-9P interface definitions)

Pin Signal Parameter
3/4/5/6/7/8/9 N/C No connect
1 Interlock Out
2 Interlock In

Short connect pin1 and pin2 make X ray source normal operation. Typical connection๏ผš

Led indicator:

ID Color Meaning
XrayOn Yellow indicate X ray on
ARC Red Arcing in oil tank
OT Red Over temperature
EP_Err Red Tube voltage error
IP_Err Red Tube current error
Power Green Power on