The HVC1206A0 series X-ray source is a compact, safe, low-leakage, and electrically stable high-voltage X-ray generator. It can operate continuously at 400W within the allowable temperature range. The X-ray source consists of a high-voltage power supply, filament power supply, X-ray tube, X-ray oil tank, and radiator. This high-frequency, self-cooling, and self-protecting X-ray generator delivers up to 120kV and 400W. It primarily includes a control box and X-ray oil tank, with RS232 interface for system control, monitoring, and firmware upgrades.
Food testing, industrial non-destructive, security application and other fields, mostly used for X-ray machines.
Item | Specification |
Input voltage | 230VACยฑ10%, 50/60Hz, 3Amps |
Output power of X ray tube | Max continuous output power 400W๏ผ120kV/3.3mA or 67kV/6mA๏ผ |
Output voltage | Rated output voltage๏ผContinuously adjustable voltage range 30kV—120kV |
Output voltage ripple๏ผยฑ0.5%๏ผpeak to peak๏ผ | |
Output voltage accuracy๏ผยฑ2% of voltage setting value | |
line regulation๏ผยฑ0.1% | |
load regulation๏ผยฑ0.1% | |
Tube current | Rated tube current๏ผContinuously adjustable current range 0.5mA-6.0mA |
Tube current accuracy๏ผยฑ0.2mA of current setting value | |
line regulation๏ผยฑ0.1% | |
load regulation๏ผยฑ0.1% | |
Rise time of output voltage | The kV rise time is <0.6 Sec at maximum power |
The kV rise time is <0.1 Sec at low voltage ( < 40kV ) | |
Filament power supply: | input voltage๏ผ24VDC |
filament voltage๏ผ2.0 to 5.3Vac | |
filament๏ผ3.0 to 3.8 Amps RMS | |
preheating time๏ผ3sec | |
Tube feature | Tube type๏ผfixed anodeใglass envelopeใtungsten target |
focus๏ผ0.8mm | |
inherent filtration๏ผ0.8mm Be๏ผ0.7mm AI | |
radiation angle๏ผ80ยบ*16ยบfan beam | |
target angle๏ผ25ยฐ | |
Cooling | transformer oil๏ผexternal circulation๏ผexternal forced air cooling |
Working temperatures | -10โ—40โ |
Storing temperature | -20โ—60โ |
System temperature protection | 60โ ยฑ3 โ of oil temperature |
Humidness | 98%๏ผNon-condensation |
Weight | Oil tank๏ผ27.3kg |
Control Unit๏ผ3.8kg | |
Installation direction | Installation in any direction |
Radiation angle | 16ยฐร80ยฐ |
X-ray leakage | less than 0.5mR/hr at 5cm from the surface of the HVC1206A0 |
JB1/AC~(AC Input Power Connector)
Pin | Signal | Parameter |
1 | L | live wire |
2 | N | Neutral line |
3 | G | PE |
JB2/COM (DMR-9S interface definitions)
Pin | Signal | Parameter |
1.4.6.7.8.9 | N/C | No connect |
2 | TXD | Data transmit |
3 | RXD | Data receive |
5 | GND | Signal gnd |
JB3/Interlock,(DMR-9P interface definitions)
Pin | Signal | Parameter |
3/4/5/6/7/8/9 | N/C | No connect |
1 | Interlock Out | |
2 | Interlock In |
Short connect pin1 and pin2 make X ray source normal operation. Typical connection๏ผ
Led indicator:
ID | Color | Meaning |
XrayOn | Yellow | indicate X ray on |
ARC | Red | Arcing in oil tank |
OT | Red | Over temperature |
EP_Err | Red | Tube voltage error |
IP_Err | Red | Tube current error |
Power | Green | Power on |
Tank size