The HVP10012A2 is a compact, portable, and reliable high-voltage X-ray generator designed for dental imaging equipment. It can continuously operate at up to 1000W for 25 seconds within the specified temperature range, relying on self-cooling without the need for additional cooling. This X-ray source can output 100kV/1000W in continuous mode (1200W in pulse mode) and features high frequency, self-cooling, self-protection, and independent control. It consists of an X-ray high-voltage oil tank and a control box, with control, status monitoring, and programming managed through a DB9 interface. Additionally, the HVP10012A2 supports both continuous and pulse operation modes.
Dental X-ray imaging equipment (CT and panoramic), head scan and mobile imaging equipment
Item | Specification |
Input voltage | 230VAC±10%, 50/60Hz,Single Phase, 6.2Amps, PF>0.98. |
Output power of X ray tube | Maximum continuous output power 1000W up to 25s (100kV, 10mA) |
Maximum pulse output power 1200W up to 40s (100kV, 12mA) | |
Output voltage | The high voltage is programmed with the range of 50 to 100KV |
Output voltage ripple:±0.5%(peak to peak) | |
Output voltage accuracy:±1% of voltage setting value | |
line regulation:±0.1% | |
load regulation:±0.1% | |
Tube current | 2mA to 10mA at 1000W MAX, continuous |
2mA to 12mA at 1200W MAX, pulse | |
Tube current accuracy:±0.1mA of current setting value | |
mA Regulation:<±0.1mA for Line Input changes of ±10%; <±0.1mA for the output voltage change over the specified range |
|
Filament power supply: | input voltage:24VDC |
filament voltage:2.5 to 6.8VAC | |
filament:3.2 to 3.8A | |
preheating time:3sec | |
Tube feature | Tube type:fixed anode、glass envelope、tungsten target |
focus:0.5mm | |
inherent filtration: 1.8mm AI | |
radiation angle: 20º, cone beam | |
target angle: 5° | |
Operation mode | Continuous mode, pulse mode |
Pulse frequency | 20-50Hz |
Pulse time | Single pulse exposure time:5.0~50ms |
Dual energy setting | Arbitrary setting of two different energy voltage levels of tube voltage within the rated range |
Max exposure time | Maximum continuous exposure time 25s at continuous mode |
Maximum continuous exposure time 40s at pulse mode | |
Kv Rise Time at maximum power: | Continuous mode: To 100kV and 10mA in less than 15ms |
Pulse mode: rise time of single pulse kV less than 1ms | |
Cooling | transformer oil,aluminum shell natural heat dissipation |
Working temperatures | -10℃—40℃ |
Storing temperature | -20℃—60℃ |
System temperature protection | 60±3℃ of oil temperature |
Humidness | 10%-95%,Non-condensation |
Weight | Oil tank:8.4kg |
Control Unit:3.6kg | |
Installation direction | Installation in any direction |
Radiation angle | cone angle of ray tapered beam 20° |
X-ray leakage | Less than 100mR/hr at 1meter from the surface of the HVP10012A2 Integrated X-Ray Source. |
JB1/AC~(AC Input Power Connector)
Pin | Signal | Parameter |
1 | L | live wire |
2 | N | Neutral line |
3 | G | PE |
JB2/COM(Communication Interface, DMR-9S(female)
Pin | Signal | Parameter |
1.4.6.7.8.9 | N/C | Undefined |
2 | TXD | Data transmit |
3 | RXD | Data receive |
5 | GND | Reference ground |
JB3 (Signal Interface, DMR-9P(male)
Pin | Signal | Parameter |
1 | Interlock Out | |
2 | Interlock In | |
3 | Syn_ Out | synchronizing signal output |
4 | Syn_ In | synchronizing signal input |
5 | GND | Reference gnd |
6 | Syn_Out | synchronizing signal output |
7 | GND | Reference ground |
8 | N/C | Undefined |
9 | Xray_ON | X-ray switch signal pin. |