The HVD801B1 ย X-ray source is a compact, low-leakage, and electrically stable high-voltage X-ray generator which operates in two modes: single-energy and dual-energy. Within the permissible operating temperature range, the X-ray source can continuously operate in single-energy mode at a maximum power of 80W for 10 seconds. In dual-energy mode, it can switch continuously between two energy levels, with adjustable voltage, current, and operating time for both levels. The X-ray source consists of a control unit, high-voltage components, and an X-ray tube. This X-ray source features high frequency, self-cooling, and high integration, with a maximum voltage of 80kV and a maximum power of 80W. It is primarily composed of a control box and a high-voltage X-ray module. The X-ray source utilizes RS232 interface for system control, monitoring, and firmware upgrades.
Mobile X-ray testing equipment in the field of human bone density.
Item | Specification |
Input voltage | 22V~26V๏ผ15Amps |
Output power of X ray tube | Max continuous output power 80W๏ผ80kV/1mA๏ผ |
Output voltage | Rated output voltage๏ผContinuously adjustable voltage range 30kV—80kV |
Output voltage ripple๏ผยฑ0.5%๏ผpeak to peak๏ผ | |
Output voltage accuracy๏ผยฑ2% of voltage setting value | |
line regulation๏ผยฑ0.1% | |
load regulation๏ผยฑ0.1% | |
Tube current | Rated tube current๏ผContinuously adjustable current range 0.1mA-1.0mA |
Tube current accuracy๏ผยฑ0.01mA of current setting value | |
line regulation๏ผยฑ0.1% | |
load regulation๏ผยฑ0.1% | |
Rise time of output voltage | Continuous mode voltage rise time: under the condition that the output voltage is less than 80kV, the output voltage rise time is less than 0.1 seconds; |
Voltage rise time in dual-energy mode: Under dual-energy operating conditions, the output voltage rise time is less than 0.1 seconds. | |
Filament power supply: | input voltage๏ผ24VDC |
filament voltage๏ผ2.0 to 5.3V | |
filament๏ผ3.0 ยฑ 0.2 Amps RMS | |
preheating time๏ผ3sec | |
Tube feature | Tube type๏ผfixed anodeใย glass envelopeย ใtungsten target |
focus๏ผ0.8mm | |
inherent filtration๏ผ1.8mm AI | |
radiation angle๏ผ40ยบcone beam | |
target angle๏ผ20ยฐ | |
Cooling | Electronic silicone solid-state packaging, natural heat dissipation. |
Working temperatures | -10โ—40โ |
Storing temperature | -20โ—60โ |
System temperature protection | 60โยฑ3โ of oil temperature |
Humidness | 98%๏ผNon-condensation |
Weight | Split type X-ray source :4kg |
Installation direction | Installation in any direction |
Radiation angle | cone angle of ray tapered beam 40ยฐ |
X-ray leakage | At 1 meter of the external surface of the tank, the radiation leakage is not more than 3mR/hr. |
JB1/DC,(DC Input Power Connector)
JB2/COM, (RJ45 interface definitions)ย
JB3,Extern Interface(12P)
Order number | Feet function |
1 | Interlock In |
2 | Interlock Out |
3 | X-ray on relay contact common |
4 | X-ray on relay contact N/C |
5 | X-ray on relay contact N/O |
6 | / |
7 | / |
8 | / |
9 | SGND |
10 | 5V |
11 | EP |
12 | IP |
Tank size
Copyright@2024, All Rights Reserved