HVP1102001 X-Ray Source

Key Features:

 

  • Compact and convenient design with highly integrated electrical modules
  • Excellent electromagnetic compatibility
  • Compatible with continuous, pulse, and dual-energy operating modes
  • Can be installed in any orientation
  • Standard digital interface for easy application and stable, reliable data exchange

Product Details

The HVP1102001 is a compact, portable, and reliable high-voltage X-ray generator designed for head imaging equipment. It can continuously operate at up to 1200W for 25 seconds within the specified temperature range, relying on self-cooling without the need for additional cooling. This X-ray source can output 110kV/1200W in continuous mode (1700W in pulse mode) and features high frequency, self-cooling, self-protection, and independent control. It consists of an X-ray high-voltage oil tank and a control box, with control, status monitoring, and programming managed through a DB9 interface. Additionally, the HVP1102001 supports both continuous and pulse operation modes.

Application:

Dental X-ray imaging equipment (CT and panoramic), head scan and mobile imaging equipment

Specification:

Item Specification
Input voltage 230VAC±10%, 50/60Hz, 9.2A,PF>0.98.
Output power of X ray tube Maximum continuous output power 1200W up to 25s (100kV, 12mA)

Maximum pulse output power 1700W up to 50s (100kV, 17mA)

Output voltage Rated output voltage:Continuously adjustable voltage range 50kV—110kV
Output voltage ripple:±0.5%(peak to peak)
Output voltage accuracy:±1% of voltage setting value
line regulation:±0.1%
load regulation:±0.1%
Tube current Rated tube current:Continuous mode: adjustable from 2 to 15mA (Max power 1200W); Pulse mode: adjustable from 2 to 20mA (maximum power 1700W).
Tube current accuracy:±0.1mA of current setting value
line regulation:±0.1mA
load regulation:±0.1mA
Filament power supply: input voltage:24VDC
filament voltage:4.0 to 8.0Vac
filament:2.2 to 4.0 Amps RMS
preheating time:3sec
Tube feature Tube type:fixed anode、glass envelope、tungsten target
focus:0.5mm
inherent filtration:1.5mm Al
radiation angle: 30º, cone beam
target angle:5°
Operation mode continuous mode, pulse mode
Pulse frequency 20-50Hz
Pulse time Single pulse exposure time:5.0~50ms
Dual energy setting Arbitrary setting of two different energy voltage levels of tube voltage within the rated range
Max exposure time Maximum continuous exposure time 25s at continuous mode
Maximum continuous exposure time 50s at pulse mode
Kv Rise Time at maximum power: Continuous mode:To 100kV and 10mA in less than 15ms
Pulse mode:rise time of single pulse kV less than 1ms
Cooling transformer oil,aluminum shell natural heat dissipation
Working temperatures -10℃—40℃
Storing temperature -20℃—60℃
System temperature protection 60±3℃ of oil temperature
Humidness 10%-95%,Non-condensation
Weight X-ray tank: 15.0kg
Control box: 3.6kg
Installation direction Installation in any direction
Radiation angle cone angle of ray tapered beam 30°
X-ray leakage Less than 100mR/hr @1meter from the surface of the HVP1102001 Integrated X-Ray Source.

JB1/AC~(AC Input Power Connector)

Pin Signal Parameter
1 L live wire
2 N Neutral line
3 G PE

JB2/COM(Communication Interface, DMR-9S(female)

Pin Signal Parameter
1.4.6.7.8.9 N/C Undefined
2 TXD Data transmit
3 RXD Data receive
5 GND Reference ground

JB3 (Signal Interface, DMR-9P(male)

Pin Signal Parameter
1 Interlock Out
2 Interlock In
3 Syn_ Out synchronizing signal output
4 Syn_ In synchronizing signal input
5 GND Reference gnd
6 Syn_Out synchronizing signal output
7 GND Reference ground
8 N/C Undefined
9 Xray_ON X-ray switch signal pin.

Short connect pin1 and pin2 make X ray source normal operation. Typical connection

Tank size

 

 

Related Products

IMG_5246
HVD801B1 X-Ray Source
IMG_5184-1
HVP702A1 Dental X-ray Machine
6
HVP1102001 X-Ray Source
IMG_5181-2
HVP10012A2 X-Ray Source